Strain Gauge analysis of non-axial loads in three-element implant-supported prostheses
DOI:
https://doi.org/10.14295/bds.2013.v16i2.866Abstract
This study evaluated “in vitro”, the microstrain around three external hexagon implants linearly placed after static loadings on non-axial points of Co-Cr superstructures. The implants were inserted into a polyurethane block and their microunit prosthetic abutments were screwed with torque of 20 Ncm. Four strain gauges were linked to the block around the implants. Onto the abutments, Co-Cr superstructures (n = 5) were screwed with torque of 10 Ncm. The static vertical loads of 30 kg were applied for 10 s during 3 repetitions in four non-axial points located perpendicularly to the long axis (A and B) and another two points at the extremity of the superstructure (A` and B`). The data were recorded with the aid of a conditioner of electrical signals and the Strain-Smart software. The results obtained were submitted to Analysis of Variance (ANOVA). It could be concluded that there were no significant differences in the effects of the applications of non-axial loads of this present study (p = 0.6464). Also, there were no differences between the displacement means, 3 mm for A and B; 5 mm for A’ and B’ (p = 0.8731).
Keywords
Biomechanics, Dental implant, Dental prosthesis.
Downloads
Published
Issue
Section
License
COPYRIGHT TRANSFER AND RESPONSIBILITY STATEMENT
(PDF | Word)
All manuscript's copyright ownership of the article "____________________________________________________________________" is transferred from the author(s) to the BRAZILIAN DENTAL SCIENCE, in the event the work is published. The manuscript has not been published elsewhere and that it has not been submitted simultaneously for publication elsewhere.
We hereby attest the study is original and does not present manipulated data, fraud or plagiarism. We made significant scientific contribution to the study and we are aware of the presented data and in agreement with the final version of the manuscript. We assume complete responsibility for the ethical aspects of the study.
This text must be printed and signed by all authors. The scanned version should be submitted as supplemental file during the submission process.