Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study

Autores

  • Carolina Enith Bullen Medina Dept. of Dentistry – Federal University of Santa Catarina – Florianópolis – SC – Brazil.
  • Caroline Freitas Rafael Federal University of Santa Catarina
  • Claudia Angela Maria Volpato Dept. of Dentistry – Federal University of Santa Catarina – Florianópolis – SC – Brazil.
  • Mutlu Özcan Dental Materials Unit – University of Zürich – Center for Dental and Oral Medicine – Clinic for Fixed and Removable Prosthodontics and Dental Materials Science – Zurich – Switzerland.
  • Diego Klee de Vasconcelos Dept. of Dentistry – Federal University of Santa Catarina – Florianópolis – SC – Brazil.

DOI:

https://doi.org/10.14295/bds.2016.v19i4.1295

Resumo

Objective: To evaluate the removal torque values on abutment screws after different torque application techniques. Materials and methods: Thirty specimens of an external hexagon implant, a prefabricated abutment and a titanium screw were distributed randomly in three groups. In group 1, the screws received a torque of 30 N.cm; group 2 received a torque of 30N.cm and a second torque after 10 min, in accordance with the technique proposed by Brending et al. and Dixon et al.; in group 3, a torque of 30 N.cm was applied and maintained for 20 s, in accordance with the technique recommended by Sella et al. The specimens were attached to a universal testing machine and a digital key was coupled to the load cell in order to control the torque value at a velocity of 1 N.cm/s. Removal torque was performed 10 minutes after torque application. Values were statistically analysed using one-way ANOVA and Tukey’s HSD test (? = 0.05). Results: The mean and standard deviations (±SD) of removal torque values found were 27.95±0.99N.cm for group 1, 28.32±0.68N.cm for group 2 and 26.89±1.03N.cm for group 3. Groups 1 and 2 exhibited statistically higher values of torque than group 3 (p ? 0.05). Conclusion: The technique recommended by Breeding et al. and Dixon et al. seemed to be the best option when considering the removal torque values of external hexagon implants.

 

Keywords: Abutment screw; Dental implants; Removal torque.

Downloads

Os dados de download ainda não estão disponíveis.

Downloads

Arquivos adicionais

Publicado

2016-11-29

Edição

Seção

Artigos de Pesquisa Clínica ou Laboratorial

Como Citar

1.
Medina CEB, Rafael CF, Volpato CAM, Özcan M, Vasconcelos DK de. Evaluation of removal torque values for titanium screws of prosthetic abutments after different torque application techniques: in vitro study. BDS [Internet]. 29º de novembro de 2016 [citado 23º de agosto de 2026];19(4):12-8. Disponível em: https://bds.ict.unesp.br/index.php/cob/article/view/1295

Artigos mais lidos pelo mesmo(s) autor(es)

<< < 1 2 3