Influence of angulation and vertical misfit in the evaluation of micro-deformations around implants

Authors

  • Vinícius Anéas Rodrigues Department of Dental Materials and Prosthodontics - Institute of Science and Technology, University State Paulista, UNESP, São José dos Campos, São Paulo
  • João Paulo Mendes Tribst Department of Dental Materials and Prosthodontics - Institute of Science and Technology, University State Paulista, UNESP, São José dos Campos, São Paulo
  • Leandro Ruivo de Santis Department of Dental Materials and Prosthodontics - Institute of Science and Technology, University State Paulista, UNESP, São José dos Campos, São Paulo
  • Dimas Renó de Lima Department of Social Dentistry, Institute of Science and Technology, University State Paulista, UNESP, São José dos Campos, São Paulo,
  • Renato Sussumu Nishioka Department of Dental Materials and Prosthodontics - Institute of Science and Technology, University State Paulista, UNESP, São José dos Campos, São Paulo

DOI:

https://doi.org/10.14295/bds.2017.v20i1.1311

Abstract

Objective: This in vitro study was to evaluate micro-strains around of implant, under the influence of angulations and vertical misfit in three-element implant-supported fixed partial dentures during axial loading by using strain gauge analysis. Materials and Methods: Three external hexagon implants with straight configuration and three external hexagon implants with angled (17°) configuration were inserted into two polyurethane blocks. To measure micro-strain, four strain gauges were bonded onto the surface of each block. Plastic copings were adapted to a standard wax pattern and cast. An axial load of 30 kgf was applied on the center of each implant for 10 seconds, using a load application device. The vertical misfit was measured at six different points by using a stereo microscope with 100-X magnification. Results: The results showed that the values for different implant angulations were significant (P= 0.0086). The Pearson’s correlation test between micro-strain and vertical misfit revealed no correlation between angled configuration (P= 0.891) and straight configuration (P= 0.568). Conclusion: The micro strain was higher for angled implants; no correlation was found between the vertical misfit and the micro strain values.

Downloads

Download data is not yet available.

Downloads

Published

2017-03-02

Issue

Section

Clinical or Laboratorial Research

How to Cite

1.
Rodrigues VA, Tribst JPM, Santis LR de, Lima DR de, Nishioka RS. Influence of angulation and vertical misfit in the evaluation of micro-deformations around implants. BDS [Internet]. 2017 Mar. 2 [cited 2026 Aug. 22];20(1):32-9. Available from: https://bds.ict.unesp.br/index.php/cob/article/view/1311